Deionized Water System LDIN-A10

Deionized Water System LDIN-A10 is smallest and most compact lab water system from tap to ultrapure water, integrating pretreatment, RO, DI, UV, UF and terminal filter into one. Its output ranges from 15 to 30 liters/hour and offers flow rate up to 2 liters/minute (with pressure tank). Features with tap water inlet, it can produce single stage RO water and deionized water. The single stage RO water’s ion rejection rate is more than 96%, and the deionized water’s resistivity is more than 15 MΩ.cm, near to 18.2 MΩ.cm. It completely meets the requirements of general chemical or biological experiments for pure water. Designed with apheliotropic LCD display, to detect the quality of deionized or ultrapure water. Equipped with Ultrafiltration module, effectively eliminate endotoxin and suitable for precise cell cultivating and IVF. Double wavelength (185 & 254 nm) ultraviolet lamp module, restrain bacteria increase and reduce TOC. It is auto self-flushing of RO membrane function, extend RO membrane life.

  • Output 30 liters/hour
    Flow rate Up to 2 liters/minute (with pressure tank)
    Pure water outlet 2 reverse osmosis water, deionized water
    Deionized Water Quality Resistivity: 16 to 18.2 M Ω cm<br/> Bacteria: N/A<br/> Particle (>0.1 µm): N/A
    RO Water Quality Ion rejection rate: 97 % to 99% (new RO membrane)<br/> Organic rejection rate: >99 %, when MW >200 Dalton<br/> Particles and bacteria rejection rate: >99 %
    Feed water requirements Tap water, Temperature 5 to 45°C, Pressure 1.0 to 4.0 Kgf/cm<sup>2</sup>
    Dimension 410 × 220 × 420 mm 
    Power Consumption 72 W
    Power Supply AC100 to 240 V, 50/60 Hz
    Weight about 18 kg
    Standard configuration Main body (Including 1 set of cartridge) +15 liters pressure tank+ TDS/conductivity test pen
    • Designed with microcomputer controlling system and electronic pressure sensor
    • Auto self-flushing of RO membrane function, extend RO membrane life
    • Equipped with top cap of pretreatment in the case can be rapidly opened to replace the cartridges without opening the case
    • Offers 4 independent ultrapure cartridges
    • Ultrapure water quality up to 18.2 MΩ.cm, with the lowest TOC level and running cost
    • Whole unit with one time injection molding process
    • Material used polypropylene, avoid rusting and keep clean, to meet GLP standard
    • Designed with apheliotropic LCD display, to detect the quality of deionized or ultrapure water
    • Offers attached portable TDS (total dissolved solid) conductivity test pen, with dry cell design, to detect the quality of feed water and RO water
    • Double wavelength (185 & 254 nm) ultraviolet lamp module, restrain bacteria´s increase and reduce TOC
    • Ultrafiltration module, effectively eliminate endotoxin and suitable for precise cell cultivating and IVF
    • Double layer PES terminal disinfection filter, assure that terminal pure water is absolutely axenic
  • Deionized Water System is used in general chemical or biological experiments to obtain deionized and RO water across Healthcare, Food & Beverage and Biochemistry microanalysis etc.

Deionized Water System LDIN-A20

  • 30 liters/hourOutput
  • Up to 2 liters/minute (with pressure tank)Flow rate
  • 2 reverse osmosis water, deionized waterPure water outlet
  • Resistivity: 16 to 18.2 M Ω cm
    Bacteria: N/A
    Particle (>0.1 µm): N/A
    Deionized Water Quality

Type III Water Purification System

It provides Type III level water which is required for chemical and physical inspection or for washing lab glassware’s

Type III Water Purification System LHWP-B11

  • ≥ 20 L / hr.Output of Pure water
  • Type IIIWater Type
  • BenchtopConfiguration
  • Double Pass RO systemProcess

Type III Water Purification System LHWP-B12

  • ≥ 30 L / hr.Output of Pure water
  • Type IIIWater Type
  • BenchtopConfiguration
  • Double Pass RO systemProcess
Type III Water Purification System LHWP-B10
  • ≥ 10 L / hr.Output of Pure water
  • Type IIIWater Type
  • BenchtopConfiguration
  • Double Pass RO systemProcess
Deionized Water System LDIN-A10
  • 30 liters/hourOutput
  • Up to 2 liters/minute (with pressure tank)Flow rate
  • 2 reverse osmosis water, deionized waterPure water outlet
  • Resistivity: 16 to 18.2 M Ω cm
    Bacteria: N/A
    Particle (>0.1 µm): N/A
    Deionized Water Quality

Deionized Water System LDIN-A20

  • 30 liters/hourOutput
  • Up to 2 liters/minute (with pressure tank)Flow rate
  • 2 reverse osmosis water, deionized waterPure water outlet
  • Resistivity: 16 to 18.2 M Ω cm
    Bacteria: N/A
    Particle (>0.1 µm): N/A
    Deionized Water Quality

Type I and Type III Water Purification System

This water purification system is a combinational system that produces Type III water after RO and Type I water after deionization in one cycle

Type I and Type III RO Water Purification System LOTW-B12

  • ≥ 30 L / hr.Output of Pure water
  • Type I and Type IIIWater Type
  • 30 LWater Tank capacity
  • BenchtopConfiguration
Type I and Type III RO Water Purification System LOTW-B10
  • ≥ 10 L / hr.Output of Pure water
  • Type I and Type IIIWater Type
  • 30 LWater Tank capacity
  • BenchtopConfiguration
Type I and Type III RO Water Purification System LOTW-B11
  • ≥ 20 L / hr.Output of Pure water
  • Type I and Type IIIWater Type
  • 30 LWater Tank capacity
  • BenchtopConfiguration

Type II Water Purification System

This water purification system provides Type II level water with consistency and reliability. It removes approximately 98% of inorganic ions and 99 % of all dissolved organic substances

Type II Water Purification System LTWP-B12

  • ≥ 30 L / hr.Output of Pure water
  • Double Pass RO systemProcess
  • Type IIWater Type
  • 30 LWater Tank capacity

Type II Water Purification System LTWP-B10

  • ≥ 10 L / hr.Output of Pure water
  • Double Pass RO systemProcess
  • Type IIWater Type
  • 30 LWater Tank capacity
Type II Water Purification System LTWP-B11
  • ≥ 20 L / hr.Output of Pure water
  • Double Pass RO systemProcess
  • Type IIWater Type
  • 30 LWater Tank capacity
Water Purification System LWPS-A11
  • Two stage(Type III and type I)Water quality
  • Tap waterTDS≤400ppm
    Water temperature: 5-40℃
    Type 3 water ≥10L/h
    Feed water
  • <5 μs/cm@25℃Conectivity
  • 18.25 MΩ.cm @ 25 ℃Resistivity

Water Purification System LWPS-C10

  • 90 liters/hourOutput 2nd stage RO water
  • Up to 2 liters/minute (when tank is full)Output ultrapure water
  • Resistivity (25°C) 18.2 M Ω cm
    TOC: < 3ppb
    Bacteria: <0.1 cfu/ml
    Particle (>0.2 µm): <1/ml
    Endotoxin: <0.001EU/ml
    RNases: <1pg/ml
    DNases: <5pg/ml
    Ultrapure Water Quality
  • Conductivity 1st stage RO water: Feed water Conductivity ×5 %
    Conductivity 2nd stage RO water: 1 to 5 µs/cm
    RO Water Quality

Water Purification System LWPS-A10

  • One stage (Type I)Water quality
  • RO water /Distilled water
    Water temperature: 5 ~ 40℃
    Feed water
  • 18.25 MΩ.cm @ 25 ℃Resistivity
  • < 5ppbTOC