Bench top Turbidity Meter LTM-B12

Bench top turbidity meter LTM-B12 with microprocessor controlled system, works on 90° scattered light principle to measure water turbidity in nephelometric turbidity unit i.e. NTU. The formazine standard solution of turbidity mentioned in ISO 7027 is used for turbidity analysis. This turbidity meter delivers highly accurate results with better numerical calculations to meet GLP standards.

  • Measuring range (NTU) 0 to 50, 0 to 100, 0 to 200, 0 to 400
    Minimum readout value (NTU) 0.01, 0.1
    Calibration point 4
    Sample volume 5 to 10 ml
    Principle method 90° scattered light
    Operating temperature 5 to 40°C
    Working temperature 5 to 40°C
    Light source Infrared-emitting diode (850 nm λ)
    Detector Silicon photodiode
    Error value ± 6 % (± 2 % F.S)
    Zero drift ± 0.5 % F.S.
    Reset functions Yes
    Memory Stores up to 20 sets data
    Output
    Dimension 400 × 300 × 250 mm
    Weight 2.5 kg
    Power batteries DC 1.5 V, 5 AA alkaline batteries
    Power supply AC 220 V, 50 Hz
    • Quick and automatic multi-points calibration
    • Back light LCD display with touch keypad
    • Wide measurement range
    • Non- linear data processing and transfer
    • Selectable automatic or manual switch mode
    • High through put stability
    • Auto power off function
    • Easy to operate
  • Used in water treatment plants, swimming pool, water recycling units, research and quality laboratories of pharmacy, life science, food, and beverage industry.

  • Accessory No. Name Quantity
    1 Specimen bottle 3 pcs
    2 400 NTU (Formazine) turbidity standard solution 1 bottle
    3 Power cable 1 pc
  • Accessory No. Name
    1 USB communication interface

Bench top Turbidity Meter (with built-in printer) LTM-C13

  • 0 to 10, 0 to 100, 0 to 200, 0 to 500, 0 to 1000Measuring range (NTU)
  • 0.01Minimum readout value (NTU)
  • 5Calibration point
  • 18 to 20 mlSample volume

Bench top Turbidity Meter LTM-B11

  • 0 to 50, 0 to 100, 0 to 200Measuring range (NTU)
  • 0.01, 0.1Minimum readout value (NTU)
  • 3Calibration point
  • 5 to 10 mlSample volume

Bench top Turbidity Meter (with built-in printer) LTM-C12

  • 0 to 10, 0 to 100, 0 to 200, 0 to 500Measuring range (NTU)
  • 0.001Minimum readout value (NTU)
  • 4Calibration point
  • 18 to 20 mlSample volume
Bench top Turbidity Meter LTM-B10
  • 0 to 100Measuring range (NTU)
  • 0.01, 0.1Minimum readout value (NTU)
  • 1Calibration point
  • 5 to 10 mlSample volume
Bench top Turbidity Meter (with built-in printer) LTM-C11
  • 0 to 100, 0 to 200, 0 to 500Measuring range (NTU)
  • 0.01Minimum readout value (NTU)
  • 3Calibration point
  • 18 to 20 mlSample volume

Portable Turbidity Meter LTM-A14

  • 0 to 100, 100 to 800Measuring range (NTU)
  • 0.01, 0.1Minimum readout value (NTU)
  • 6Calibration point
  • 5 to 10 mlSample volume

Portable Turbidity Meter LTM-A12

  • 0 to 100, 100 to 400Measuring range (NTU)
  • 0.01, 0.1Minimum readout value (NTU)
  • 5Calibration point
  • 5 to 10 mlSample volume

Portable turbidity meter LTM-A16

  • LED/860 nm (Wavelength)Light source
  • LCDReadings display
  • NTU, FTU, EBC, ASBCUnit switches
  • 0.00-20.0; 20.01-200; 200.01-1000 NTUMeasurement range
Bench top Turbidity Meter (with built-in printer) LTM-C14
  • 0 to 10, 0 to 100, 0 to 200, 0 to 500, 0 to 1000Measuring range (NTU)
  • 0.001Minimum readout value (NTU)
  • 5Calibration point
  • 18 to 20 mlSample volume
Bench top Turbidity Meter (with built-in printer) LTM-C15
  • 0 to 10, 0 to 100Measuring range (NTU)
  • 0.001Minimum readout value (NTU)
  • 2Calibration point
  • 18 to 20 mlSample volume

Bench top Turbidity Meter LTM-B14

  • 0 to 50, 0 to 100, 0 to 200, 0 to 400, 0 to 800, 0 to 1000Measuring range (NTU)
  • 0.01, 0.1Minimum readout value (NTU)
  • 6Calibration point
  • 5 to 10 mlSample volume

Portable Turbidity Meter LTM-A11

  • 0 to 100, 0 to 200Measuring range (NTU)
  • 0.01, 0.1Minimum readout value (NTU)
  • 4Calibration point
  • 5 to 10 mlSample volume

Portable Turbidity Meter LTM-A15

  • 0 to 100, 100 to 1000Measuring range (NTU)
  • 0.01, 0.1Minimum readout value (NTU)
  • 7Calibration point
  • 5 to 10 mlSample volume
Portable Turbidity Meter LTM-A10
  • 0 to 200Measuring range (NTU)
  • 0.1Minimum readout value (NTU)
  • NilCalibration point
  • 5 to 10 mlSample volume
Bench top Turbidity Meter (with built-in printer) LTM-C10
  • 0 to 100, 0 to 200Measuring range (NTU)
  • 0.01Minimum readout value (NTU)
  • 2Calibration point
  • 18 to 20 mlSample volume

pH Meters (Portable)

Labtron portable meters are used in quality analysis of water and other solutions. With portability, waterproof features, finds applications in indoor as well as outdoor measurement of pH, ORP, dissolved oxygen, water hardness, salinity, TDS etc.

Portable pH meter LPRPM-A10

  • (-2.00) to 16.00 pHpH range
  • ± 0.01 pHpH accuracy
  • 0.01 pHpH resolution
  • 2 pointsCalibration points

Portable pH/ORP meter LPRPM-A11

  • (-2.00) to 20.00 pHpH range
  • ± 0.002 pHpH accuracy
  • 0.1, 0.01, 0.001 pHpH resolution
  • 1 - 5 pointsCalibration points

ORP Tester

Labtron pocket testers are small and compact in size, gives fast and accurate results. The Pocket testers consist of pH, ORP, TDS, dissolved oxygen, salinity and conductivity tester for various laboratory as well as outdoor applications. Measurements can be taken by inserting the probe in to the test solution and after gentle stirring, stable readings can be noted down.

Pocket ORP tester LPOT-A11
  • BNCConnector
  • (-999) to 999 mVORP measurement range
  • ± 2 mVORP accuracy
  • 1 mVORP resolution
Pocket ORP tester LPOT-A10
  • Platinum sensorSensor type
  • 6 pinConnector
  • (-999) to 999 mVORP measurement range
  • ± 2 mVORP accuracy

Online pH Controller LOPC-A10

  • 01.00 to 15.00 pHpH measurement range
  • 1 to 3 pointspH calibration points
  • USA (pH 4.01, 7.00, 10.01) or NIST (pH 4.01, 6.86, 9.18)pH Buffer Solutions
  • ± 0.01 pHpH accuracy

Benchtop Conductivity Meter

Labtron benchtop meters are easy to use. The LCD display, manual or automatic mode of calibration, universal connector and other features are included in this range to produce accurate, reliable results. It consist of benchtop meter for measuring pH, conductivity, dissolved oxygen content, Ion concentration, water hardness level. These meters are suitable for simple to complex applications in water quality testing, research and analysis etc.

Benchtop Conductivity Meter LBCM-A13

  • 0.000 μS/cm to 3000mS/cmConductivity measurement range
  • 5.00 Ω·cm to 100.0 MΩ·cmResistivity measurement range
  • 0.000 mg/L to 1000g/LTDS measurement range
  • 0.00 to 8.00 %Salinity measurement range
Benchtop Conductivity Meter LBCM-A14
  • {"Conductivity measurement range":"0.01-20.0, 200.0, 2000 \u00b5S\/cm, 20.00, 200.0 mS\/cm","Conductivity accuracy":"\u00b11% F.S.","Conductivity resolution":"0.001, 0.01, 0.1, 1","Conductivity calibration points":"1 to 3 points","Conductivity calibration solutions":"10 \u00b5S\/cm , 84 \u00b5S\/cm, 1413 \u00b5S\/cm, 12.88 mS\/cm, 111.8 mS\/cm"}Conductivity Mode
  • {"TDS measurement Range":"0 to 10.00, 100.0, 1000 ppm, 10.00, 100 ppt (Max. 200 ppt)","TDS Accuracy":"\u00b1 1 % F.S.","TDS Factor":"0.01 to 1.00 (default 0.5)","TDS Calibration points":"3 points"}TDS Mode
  • {"Temperature range":"0 to 105\u00b0C, 32 to 221\u00b0F","Temperature accuracy":"\u00b11\u00b0C","Temperature resolution":"0.1\u00b0C, 0.1\u00b0F","Temperature calibration points":"1 point","Temperature Compensation Range and Mode":"Manual or Automatic, 0 to 100\u00b0C , 32 to 212\u00b0F","Temperature coefficient":"0.0 to 10.0 % per \u00b0C"}Temperature Mode
  • K=0.1, 1, 10Cell constant
Benchtop conductivity meter LBCM-A11
  • 0 - 20.0, 200.0, 2000 µS / cm, 200.0 mS / cmConductivity measurement range
  • ± 1 % F.S.Conductivity accuracy
  • 0.01, 0.1, 1Conductivity resolution
  • 10 µS / cm , 84 µS / cm, 1413 µS / cm, 12.88 mS / cm, 111.8 mS / cmConductivity calibration solutions

Benchtop Multi-parameter Water Quality Meter LMPWM-A30

  • pH range -2.00 to 20.00 pH
    pH Accuracy ± 0.002 pH
    pH Resolution 0.1/0.01/0.001 pH, selectable
    pH Calibration Points 1 to 5 Points
    pH Buffer Solutions USA (pH 4.0/7.00/10.01) or NIST (pH 4.01/6.86/9.18) or DIN buffers
    pH Mode
  • mV Range -1999.9 to 1999.9 mV
    mV Accuracy ± 0.2 mV
    mV Resolution 0.1/1 mV, selectable
    mV Calibration Points 1 Point
    mV Mode
  • Conductivity Range 0.01 to 20.00, 200.0, 2000 µS/cm, 20.00, 200.0 mS/cm
    Conductivity Accuracy ± 0.5% F.S.
    Conductivity Resolution 0.01/0.1/1
    Conductivity Calibration Points 1 to 5 Points
    Conductivity Solutions 10, 84, 1413 µS/cm, 12.88, 111.8 mS/ cm
    Conductivity Temperature coefficient 0.0 to 10.0 %/°C
    Compensation mode Linear or Pure Water
    Cell Constant K=0.1, 1, 10 or Custom
    Reference temperature 20 or 25°C
    Conductivity Mode
  • TDS Range 0 to 10.00, 100.0, 1000 ppm, 10.00, 100 ppt (Max. 200 ppt)
    TDS Accuracy ± 1 % F.S.
    TDS Factor 0.1  to 1.0 (default 0.5)
    TDS Mode

Benchtop Multi-parameter Water Quality Meter LMPWM-A32

  • pH range -2.00 to 20.00 pH
    pH Accuracy ± 0.002 pH
    pH Resolution 0.1/0.01/0.001 pH, selectable
    pH Calibration Points 1 to 5 Points
    pH Buffer Solutions USA (pH 4.0/7.00/10.01) or NIST (pH 4.01/6.86/9.18) or DIN buffers
    pH Mode
  • mV Range -1999.9 to 1999.9 mV
    mV Accuracy ± 0.2 mV
    mV Resolution 0.1/1 mV, selectable
    mV Calibration Points 1 Point
    mV Mode
  • Conductivity Range 0.01 to 20.00, 200.0, 2000 µS/cm, 20.00, 200.0 mS/cm
    Conductivity Accuracy ± 0.5% F.S.
    Conductivity Resolution 0.01/0.1/1
    Conductivity Calibration Points 1 to 5 Points
    Conductivity Solutions 10, 84, 1413 µS/cm, 12.88, 111.8 mS/ cm
    Conductivity Temperature coefficient 0.0 to 10.0 %/°C
    Compensation mode Linear or Pure Water
    Cell Constant K=0.1, 1, 10 or Custom
    Reference temperature 20 or 25°C
    Conductivity Mode
  • Dissolved Oxygen Range 0.00 to 20.00 mg/ L
    Dissolved Oxygen Accuracy ± 0.2 mg/ L
    Dissolved Oxygen Resolution 0.1/ 0.01 mg/ L, selectable
    Dissolved Oxygen Calibration Points 1 or 2 Points
    % Saturation Range 0.0 to 200.0 % with ± 2 % accuracy
    Pressure correction 60 to 112.5 kPa, 450 to 850 mmHg
    Salinity correction 0 to 50 g/ L
    Dissolved Oxygen Mode
Benchtop Multi-parameter Water Quality Meter LMPWM-A31
  • pH range -2.00 to 20.00 pH
    pH Accuracy ± 0.002 pH
    pH Resolution 0.1/0.01/0.001 pH, selectable
    pH Calibration Points 1 to 5 Points
    pH Buffer Solutions USA (pH 4.0/7.00/10.01) or NIST (pH 4.01/6.86/9.18) or DIN buffers
    pH Mode
  • mV Range -1999.9 to 1999.9 mV
    mV Accuracy ± 0.2 mV
    mV Resolution 0.1/1 mV, selectable
    mV Calibration Points 1 Point
    mV Mode
  • Conductivity Range 0.01 to 20.00, 200.0, 2000 µS/cm, 20.00, 200.0 mS/cm
    Conductivity Accuracy ± 0.5% F.S.
    Conductivity Resolution 0.01/0.1/1
    Conductivity Calibration Points 1 to 5 Points
    Conductivity Solutions 10, 84, 1413 µS/cm, 12.88, 111.8 mS/ cm
    Conductivity Temperature coefficient 0.0 to 10.0 %/°C
    Compensation mode Linear or Pure Water
    Cell Constant K=0.1, 1, 10 or Custom
    Reference temperature 20 or 25°C
    Conductivity Mode
  • TDS Range 0 to 10.00, 100.0, 1000 ppm, 10.00, 100 ppt (Max. 200 ppt)
    TDS Accuracy ± 1 % F.S.
    TDS Factor 0.1  to 1.0 (default 0.5)
    TDS Mode