Atomic Absorption Spectrophotometer LAAS-A22

Atomic Absorption spectrophotometer comes with flame, graphite furnace and hydride generation system to meet different customer demand. It is provided with deuterium hollow cathode lamp to eliminate the interference of molecular absorption and Titanium Burner to improve the analysis efficiency and accuracy of the flame. The autosampler has integrated graphite furnace design with high-precision syringes to enter volume as minimum as 0.5μl with intelligent online dilution and concentration function. It has great applications research and medicine, environmental science and other industries.

  • Monochromator
    Wavelength range 180 – 900 nm
    Spectral Bandwidth 0.1 nm,0.2 nm,0.4 nm, 1.0 nm, 2.0 nm automatic setting
    Wavelength accuracy ≤ 0.15 nm
    Wavelength repeatability ± 0.1 nm
    Baseline drift ≤ ± 0.002 Abs/30 min (Static) ≤ ± 0.005 Abs/30 min (Dynamic)
    Grating groove density
    Sample Volume 1 - 100 μl
    The volume of sample better than 1 %(at the time of 10 mL - 100 mL)
    Repeated Sampling Frequency up to 99 Times
    Cleaning and waste container volume Each 500 mL
    Absorbance Range 0 - 4 A
    The Program Automatic temperature control up to 20 order
    Temperature Range room temperature to 3000 ℃
    Detector imported photomultiplier tube
    Heating Mode vertical heating
    Temperature Control Method Vertical optical temperature monitoring graphite tube wall temperature
    Temperature Control The furnace enriched up to 20 times
    Resolution
    Lamp holder ≤ 6 lamps automatic turret, automatic alignment
    Lamp current adjustment range
    Lamp power supply mode pulsed power supply
    Optical System Large 1800 /mm grating ruling, full closed optical system
    Atomization Chamber Polymer explosion-proof spray chamber
    Characteristics Volume 0.5 × 10<sup>-12</sup>g (Cd)
    Characteristic concentration (Cu) 0.015 μg / mL / 1 %
    Detection Limit 0.4 × 10<sup>-12</sup>g (Cd)
    Detection Limit (Cu) 0.002 μg/mL
    Background calibration ability
    Burner Full titanium combustion head, 50 mm or 100 mm general combustion head
    Nebulizer Atomizer efficient glass atomizer,can also be customized
    Precision RSD ≤ 0.5
    Ignition Type Microcomputer control, automatic ignition
    Sprayer
    Room Spray
    Sample Tray 70 sample cups, 6 reagent cup
    The Smallest Lincrement 0.1 μl
    Safety measures The graphite tube damage, water flow Air pressure and other alarm temperature overheating protection
    Gas Control Automatic gas control system
    Measurement Flame method, Hydride generation-Atomic absorption method
    Concentration calculation Standard curve method(1 - 3 times curve), automatic fitting
    The Cooling Water Can choose cooling water circulation system
    The number of repeated measurements 1 to 99, automatic calculation of average concentration, standard deviation, relative standard deviation.
    Print results Print the final stages of test data or analysis. The results can be exported into Excel worksheet.
    Software Environment
    Operating temperature
    Working humidity
    Power supply 110 / 220 V (+5 % ~ (-10) %), 60 / 50 Hz; 5000 VA
    Power (VAC)
    Gross Weight
    Weight
    Dimension (L × W × H)
    • High precision automatic optical system
    • Graphite furnace temperature control
    • High technical indicators
    • Deuterium lamp for background correction system
    • Polymer automization chamber
    • Powerful, intelligent user-friendly design
    • Automatic analysis & Sampler
  • Atomic Absorption spectrophotometer is used in scientific research, quality control, disease control, environmental protection, metallurgy, agriculture, chemical industry, and analysis of samples, equipment maintenance, simple and convenient.

Related Products of Atomic Absorption Spectrophotometer

Labtron Atomic Absorption Spectrophotometer designed for accurate elemental analysis. Each of them features integrated flame and graphite furnace atomization systems, automatic wavelength scanning, and peak searching capabilities. Safety alarms ensure secure operation. Our Atomic Absorption Spectrophotometer delivers reliable results across a wide range of wavelengths with precise optics and temperature control.

Atomic Absorption Spectrophotometer LAAS-A21

  • 180 – 900 nmWavelength range
  • 0 nm, 0.2 nm, 0.4 nm, 1.0 nm, 2.0 nm (5 steps with automatic changeover)Spectral Bandwidth
  • ≤ 0.15 nmWavelength accuracy
  • ± 0.1 nmWavelength repeatability

Atomic Absorption Spectrophotometer LAAS-A20

  • 180 - 900 nmWavelength range
  • 0, 0.2, 0.4, 1.0, 2.0nm (Automatic setting)Spectral Bandwidth
  • ≤ 0.15 nmWavelength accuracy
  • ± 0.1 nmWavelength repeatability

Atomic Absorption Spectrophotometer LAAS-A13

  • 190 to 900 nmWorking spectral range
  • 0.1, 0.2, 0.4, 1.0 and 2.0 nmSpectral Bandwith
  • ± 0.15 nmWavelength accuracy
  • ≤ 0.04 nmWavelength repeatability
Atomic Absorption Spectrophotometer LAAS-A11
  • 190 - 900 nmWavelength range
  • 0.2 nm, 0.4 nm, 0.7 nm, 1.4 nm, 2.4 nm, 5.0 nmSpectral Bandwidth
  • ≤ ± 0.5 nmWavelength accuracy
  • ≤ 0.3 nm(single direction)Wavelength repeatability
Flame Atomic Absorption Spectrophotometer LAAS-A12
  • 190 nm ~ 900 nmWavelength range
  • 0.1 nm, 0.2 nm, 1.0 nm, 2.0 nmSpectral bandwidth
  • ± 0.25 nmWavelength accuracy
  • 0.15 nmWavelength repeatability

Double Beam UV/Vis Spectrophotometer

Labtron Double Beam UV/Vis Spectrophotometer offer automatic wavelength adjustments with their silicon photodiode detectors. Their dual beam designs cover a broad wavelength range. They support multiple measurement methods to meet diverse analytical requirements. Our spectrophotometer ensures precise results and ease of use with auto calibration, user-friendly interfaces, and advanced optics.

Double Beam UV/Vis Spectrophotometer LUS-B31

  • 190 to 1100 nmWavelength range
  • 1 nmSpectral bandwidth
  • ± 0.1nm @ 656.1 nm, ± 0.3 nm @ allWavelength Accuracy
  • ≤ 0.1 nmWavelength Repeatability

Double Beam UV/Vis Spectrophotometer LUS-B30

  • 190 to 1100 nmWavelength range
  • 2 nmSpectral bandwidth
  • ± 0.1nm @ 656.1 nm, ± 0.3 nm @ allWavelength Accuracy
  • ≤ 0.1 nmWavelength Repeatability
Double Beam UV/Vis Spectrophotometer LUS-B32
  • 190 to 1100 nmWavelength range
  • 0.5/1/2/4/5 nmSpectral bandwidth
  • ± 0.1nm @ 656.1 nm, ± 0.3 nm @ allWavelength Accuracy
  • ≤ 0.1 nmWavelength Repeatability

Fluorescence Spectrophotometer

Labtron Fluorescence Spectrophotometer are built with sensitive signal detection technology and advanced optical systems. They offer impressive fluorescence detection limits and exceptional signal-to-noise ratios. With a dual beam ratio excitation light monitoring system, they ensure stable fluorescence signals. Our Fluorescence Spectrophotometer have built-in optical gates, which enable effective analysis of unstable samples.

Fluorescence Spectrophotometer LFS-A11
  • 150 W Xenon LampExcitation Source
  • 200 nm to 900 nmExcitation Wavelength
  • 1 nm, 2 nm, 5 nm, 10 nm, 20 nmExcitation Slit
  • 1 nm, 2 nm, 5 nm, 10 nm, 20 nmEmission Slit

Fluorescence Spectrophotometer LFS-A10

  • 150 W Xenon lampLight source
  • 200 nm to 900 nmExcitation wavelength
  • 200 nm to 900 nmEmission wavelength
  • 2 nm, 5 nm, 10 nm, 20 nmExcitation slit

Fluorescence Spectrophotometer LFS-A12

  • 150 W Xenon lampLight source
  • 200 nm to 900 nmExcitation wavelength
  • 200 nm to 900 nmEmission wavelength
  • 2 nm, 5 nm, 10 nm, 20 nmExcitation slit

Single Beam UV/Visible Spectrophotometer

Labtron Single Beam UV/Visible Spectrophotometer feature a wide range of wavelengths and multiple options for spectral bandwidth selection. They facilitate quantitative analysis and photometric measurements through intuitive interfaces. They offer flexibility with various installed light sources. Our Single Beam UV/Vis Spectrophotometer ensure high performance by incorporating automatic wavelength adjustment.

Single Beam UV/Vis Spectrophotometer LUS-A50
  • 190 to 1100 nmWavelength range
  • 2 nmSpectral bandwidth
  • ± 0.1nm @ 656.1 nm, ± 0.3 nm @ allWavelength Accuracy
  • ≤ 0.1 nmWavelength Repeatability
Single Beam UV/Vis Spectrophotometer LUS-A52
  • 190 to 1100 nmWavelength range
  • 0.5/1/2/4/5 nmSpectral bandwidth
  • ± 0.1nm @ 656.1 nm, ± 0.3 nm @ allWavelength Accuracy
  • ≤ 0.1 nmWavelength Repeatability

Single Beam UV/Vis Spectrophotometer LUS-A51

  • 190 to 1100 nmWavelength range
  • 1 nmSpectral bandwidth
  • ± 0.1nm @ 656.1 nm, ± 0.3 nm @ allWavelength Accuracy
  • ≤ 0.1 nmWavelength Repeatability

Table Top Spectrophotometer

Labtron Table Top Spectrophotometer features ETC technology ensuring real-time calibration for precise results. They support multiple measurement apertures, including options for small, medium, and large sample sizes. They also have specialized configurations for liquids and reflective materials. Their high-capacity rechargeable batteries provide long-lasting performance. Our Table Top Spectrophotometer offer a wide range of color spaces and color difference formulas catering to diverse application needs.

Table Top Spectrophotometer LTS-A16

  • 360 nm to 780 nmWavelength Range
  • 10 nmWavelength Pitch
  • Silicon Photodiode ArraySensor
  • Concave GratingGrating Method
Table Top Spectrophotometer LTS-A15
  • 400 nm to 700 nmWavelength Range
  • 10 nmWavelength Interval
  • High sensitivity silicon photodiodeSensor
  • LEDLight Source
Table top spectrophotometer LTS-A10
  • d/8 (Diffused lighting, 8° observation angle)
    SCS optical engine (light splitting and integration system)
    ETC (real time calibration technology )
    SCI (specular reflection included) & SCE(specular reflection excluded) simultaneous measurement
    Illumination system
  • 40 mm, alvan diffused reflection surface coatingSize of integrated sphere
  • CLEDIllumination light source
  • Dual light path sensor arraySensor

Nano Spectrophotometer

Labtron Nano Spectrophotometer enables rapid analysis of small sample volumes and features a wide wavelength range for precise measurements. They ensure high repeatability in detecting micro-volume quantities of nucleic acids and other components. They include built-in computers with USB connections and user-friendly interfaces. Our Nano Spectrophotometer are Ideal for concentrated samples without the need for dilution or baseline adjustments.

Nano Spectrophotometer LNS-A21

  • 200 - 850 nmWavelength range
  • 1 mm, 2 mm, 5 mm, 10 mmPath length
  • ± 1 nmWavelength accuracy
  • 2 nm (FWHM at Hg 546 nm)Wavelength Resolution

Nano Spectrophotometer LNS-B11

  • 230 nm, 260 nm, 280 nmWavelength range
  • 1.0 mm, 0.2 mmPath length
  • 0.3 - 2.0 µlMinimum Sample size
  • 0.2 - 75A (10 mm equivalent absorbance)Absorbance range
Nano Spectrophotometer LNS-A10
  • 200 - 800 nmWavelength range
  • 0.2 mm (For high concentration measurement); 1.0 mm (For ordinary)Path length
  • ± 1 nmWavelength accuracy
  • ≤ 3 nm (FWHM at Hg 546 nm)Wavelength Resolution