Fluorescence Spectrophotometer LFS-A10

Fluorescence Spectrophotometer LFS-A10 consist of a high quality 150 W Xenon light source and photoelectric multiplier tube detectors for providing sufficient light intensity signal to detect sensitivity which ranges between 200 nm to 900 nm. Features fluorescence detection limit of ≤ 5×10-11 g/ml and Water Raman peak of S/N ≥ 200. It consists of the following characteristics, namely, high detection sensitivity, fast scanning speed, high dynamic range, fast 3D scanning, etc. The main principles includes measurement of absorption and fluorescence. Absorbance measurements can be taken via one or more wavelengths. The fluorescence unit increases the measuring range by 1000 for detecting DNA.

  • Light source 150 W Xenon lamp
    Excitation wavelength 200 nm to 900 nm
    Emission wavelength 200 nm to 900 nm
    Excitation slit 2 nm, 5 nm, 10 nm, 20 nm
    Emission slit 2 nm, 5 nm, 10 nm, 20 nm
    Wavelength accuracy ±0.4 nm
    Wavelength repeatability ≤ 0.2 nm
    Signal-to-Noise Ratio Raman peak of water (P-P): S/N ≥ 200 (10 nm Slit)
    Detection limit ≤5×10-11 g/ml (Quinine Sulphate Solution)
    Linearity γ ≥ 0.995
    Peak repeatability ≤ 1.5%
    Stability (10 min) Zero Drift ±0.3
    Stability (10 min) Value Limit ±1.5%
    Wavelength scan speed Multi-speed level, Maximum at 48000 nm/min
    Photometric value 0.00 to 10000.00
    Data interface USB 2.0
    Maximum power consumption 200 W
    Power Source AC 220 V/50 Hz; 110 V/60 Hz
    Dimension 380×445×310 mm
    Net weight 12 kg
    Gross weight 14 kg
    • High sensitivity is fetched from high efficiency optical design and weak signal detection technology
    • Excitation and emission wavelength range from 200 nm to 900 nm to meet most of the fluorescence analysis
    • Fluorescence detection limit of ≤ 5×10-11 g/ml and Water Raman peak of S/N ≥ 200
    • Classic fluorescence spectra and high quality 3D fluorescence spectra can be achieved via high scanning speed at 48000 nm/min
    • Wide Spectral measurement range is achieved using a double monochromator
    • Excitation light monitoring system consist of a light dual beam ratio to ensure the fluorescence signal is high and stable
    • Unstable samples is analysed via built-in optical gate
    • It is a highly advance structure with fine finishing
  • Fluorescence Spectrophotometer finds application in material research, pharmaceutical analysis, biochemical and clinical testing, water quality analysis and control, food safety testing.

  • Accessories No. Accessories Name Functions
    1 Membrane sample rack For membrane samples
    2 Powder sample rack For powder samples
    3 Up conversion fluorescent holder Base holder for other racks
    4 Octave Double-frequency elimination filter (300 nm,350 nm,400 nm,450 nm,500 nm,550 nm,600 nm,650 nm for one set) Remove the interference of double-frequency peaks of these wavelengths
    5 Auto polarization filter -
    6 Sample cell jacket -
    7 Peltier temperature (15 to 40°C) control holder -

Related Products of Fluorescence Spectrophotometer

Labtron Fluorescence Spectrophotometer are built with sensitive signal detection technology and advanced optical systems. They offer impressive fluorescence detection limits and exceptional signal-to-noise ratios. With a dual beam ratio excitation light monitoring system, they ensure stable fluorescence signals. Our Fluorescence Spectrophotometer have built-in optical gates, which enable effective analysis of unstable samples.

Fluorescence Spectrophotometer LFS-A11

  • 150 W Xenon LampExcitation Source
  • 200 nm to 900 nmExcitation Wavelength
  • 1 nm, 2 nm, 5 nm, 10 nm, 20 nmExcitation Slit
  • 1 nm, 2 nm, 5 nm, 10 nm, 20 nmEmission Slit

Fluorescence Spectrophotometer LFS-A12

  • 150 W Xenon lampLight source
  • 200 nm to 900 nmExcitation wavelength
  • 200 nm to 900 nmEmission wavelength
  • 2 nm, 5 nm, 10 nm, 20 nmExcitation slit

Visible Spectrophotometer

Labtron Visible Spectrophotometer feature high-resolution grating systems that cover a broad wavelength range. These instruments offer multiple modes, including transmittance, absorbance, concentration, and autocalibration. They utilize adaptable light sources and silicon photodiode detectors for precise measurements. Our visible spectrophotometers provide exceptional stability and precision, making them ideal for industrial applications and research.

Vis Spectrophotometer LVS-A10

  • 330 to 1000 nmWavelength Range
  • 5 nmSpectral Bandwidth
  • ± 2 nmWavelength Accuracy
  • ≤ 1.0 nmWavelength Repeatability

Vis Spectrophotometer LVS-A11

  • 335 ~ 1000 nmWavelength Range
  • 4 nmSpectral Bandwidth
  • ± 2 nmWavelength Accuracy
  • 1 nmWavelength Repeatability
Vis Spectrophotometer LVS-B10
  • 325 ~ 1000 nmWavelength Range
  • AutomaticSetting Wavelength
  • 2 nmSpectral Bandwidth
  • ± 1.0 nmWavelength Accuracy

Xenon Lamp Spectrophotometer

Labtron Xenon Lamp Spectrophotometer comes with a high-quality flashing Xenon lamp that requires no preheating and has a lifespan of 3 years. Their multiple work modes, including T, A, C, and E, offer exceptional flexibility for various types of measurements. They use imported silicon photodiode detectors for high sensitivity and stability. Our Xenon Lamp Spectrophotometer’s large sample chamber supports the attachment of optional accessories, enhancing adaptability.

Xenon Lamp Spectrophotometer LXLS-A10
  • Double Beam, Grating 1200 lines/mmOptical system
  • 190 nm to 1100 nmWavelength Range
  • 2 nmSpectral Bandwidth
  • ± 0.1 nm @ 656.1 nm, ± 0.3 nm @ allWavelength Accuracy

Xenon Lamp Spectrophotometer LXLS-A11

  • Double Beam, Grating 1200 lines/mmOptical system
  • 190 nm to 1100 nmWavelength Range
  • 1 nmSpectral Bandwidth
  • ± 0.1 nm @ 656.1 nm, ± 0.3 nm @ allWavelength Accuracy

Xenon Lamp Spectrophotometer LXLS-A13

  • Double Beam, Grating 1200 lines/mmOptical system
  • 190 nm to 1100 nmWavelength Range
  • 2 nmSpectral Bandwidth
  • ± 0.1 nm @ 656.1 nm, ± 0.3 nm @ allWavelength Accuracy

Nano Spectrophotometer

Labtron Nano Spectrophotometer enables rapid analysis of small sample volumes and features a wide wavelength range for precise measurements. They ensure high repeatability in detecting micro-volume quantities of nucleic acids and other components. They include built-in computers with USB connections and user-friendly interfaces. Our Nano Spectrophotometer are Ideal for concentrated samples without the need for dilution or baseline adjustments.

Nano Spectrophotometer LNS-A21
  • 200 - 850 nmWavelength range
  • 1 mm, 2 mm, 5 mm, 10 mmPath length
  • ± 1 nmWavelength accuracy
  • 2 nm (FWHM at Hg 546 nm)Wavelength Resolution
Nano Spectrophotometer LNS-B11
  • 230 nm, 260 nm, 280 nmWavelength range
  • 1.0 mm, 0.2 mmPath length
  • 0.3 - 2.0 µlMinimum Sample size
  • 0.2 - 75A (10 mm equivalent absorbance)Absorbance range

Nano Spectrophotometer LNS-A10

  • 200 - 800 nmWavelength range
  • 0.2 mm (For high concentration measurement); 1.0 mm (For ordinary)Path length
  • ± 1 nmWavelength accuracy
  • ≤ 3 nm (FWHM at Hg 546 nm)Wavelength Resolution

Fluorescence Spectrophotometer

Labtron Fluorescence Spectrophotometer are built with sensitive signal detection technology and advanced optical systems. They offer impressive fluorescence detection limits and exceptional signal-to-noise ratios. With a dual beam ratio excitation light monitoring system, they ensure stable fluorescence signals. Our Fluorescence Spectrophotometer have built-in optical gates, which enable effective analysis of unstable samples.

Fluorescence Spectrophotometer LFS-A10

  • 150 W Xenon lampLight source
  • 200 nm to 900 nmExcitation wavelength
  • 200 nm to 900 nmEmission wavelength
  • 2 nm, 5 nm, 10 nm, 20 nmExcitation slit
Fluorescence Spectrophotometer LFS-A11
  • 150 W Xenon LampExcitation Source
  • 200 nm to 900 nmExcitation Wavelength
  • 1 nm, 2 nm, 5 nm, 10 nm, 20 nmExcitation Slit
  • 1 nm, 2 nm, 5 nm, 10 nm, 20 nmEmission Slit
Fluorescence Spectrophotometer LFS-A12
  • 150 W Xenon lampLight source
  • 200 nm to 900 nmExcitation wavelength
  • 200 nm to 900 nmEmission wavelength
  • 2 nm, 5 nm, 10 nm, 20 nmExcitation slit

Portable Spectrophotometer

Labtron Portable Spectrophotometer utilize traditional whiteboards in their optical systems to ensure accurate measurements. They incorporate ETC technology for precise calibration. Include automated gloss compensating technology for accurate color measurements. Their extensive storage capacity provides long-term data retention. Our Portable Spectrophotometer are equipped with a TFT touch screen, a built-in camera, and high-capacity rechargeable batteries.

Portable Spectrophotometer LSP-A30

  • 45/0 (45 ring shaped illumination,0° observation angle)Illumination
  • Light splitting reflectivity:standard deviation within 0.08% Color values: ΔE*ab<=0.03 (After calibration, standard deviation of 30 measurements on test white board, 5 second intervals) Maximum:0.05Repeatability
  • High sensitivity silicon photodiodeSensor
  • -Test angle

Portable spectrophotometer LSP-A10

  • d/8 (Diffused lighting, 8° observation angle)
    SCS optical engine (light splitting and integration system)
    ETC (real time calibration technology )
    SCI (specular reflection included) & SCE(specular reflection excluded) simultaneous measurement
    Illumination
  • Light splitting reflectivity:standard deviation within 0.08%
    Color values:ΔE*ab<=0.03(After calibration, standard deviation of 30 measurements on test white board, 5 second intervals)
    Maximum:0.05
    Repeatability
  • Blue light enhanced sensor arraySensor
  • 40 mm, alvan diffused reflection surface coatingSize of integrated sphere
Portable Spectrophotometer LSP-A31
  • 45/0 (45 ring shaped illumination,0° observation angle)Illumination
  • Light splitting reflectivity:standard deviation within 0.08% Color values: ΔE*ab<=0.03 (After calibration, standard deviation of 30 measurements on test white board, 5 second intervals) Maximum:0.05Repeatability
  • High sensitivity silicon photodiodeSensor
  • 60°Test angle