HPLC system LHLC-B10

This HPLC system is incorporated with advanced high pressure injection technique. The detector is equipped with 24 bits AD transducer to assure low noise, low drift and high sensibility. This HPLC system can conveniently change the sample input volume. It comes with dual piston reciprocating and series pump for consistent and stable flow.

  • High pressure pump High pressure gradient pump
    Peak operating pressure 45 MPa, Stainless steel pump
    Pressure ± 0.02 MPa
    Flow rate range 0.001 - 9.999 mL / min
    Flow rate precision ± 5 %
    Flow rate repeatability RSD ≤ 0.1 %
    Power 220V ± 10 %, 50 Hz, 250 W
    Size (W × H × D) 520 × 420 × 160 mm
    UV Detector
    • Wavelength: 190 - 680 nm
    • Wavelength error: ≤ ± 1 nm
    • Baseline noise: ≤ ± 2 × 10<sup>-5</sup> AU (Methyl alcohol, 1 mL / min, 254 nm, 20 ºC)
    • Wavelength repeatability: 0.2 nm
    • Flow cell: 8 µL
    • Drift: ≤ 5 × 10<sup>-4</sup> AU/h (Methyl alcohol, 1 mL / min, 254 nm, 20 ºC)
    • Limit of detection: ≤ 2 × 10<sup>-9</sup> g / mL (Naphthalene methyl alcohol solvent)
    Automatic Injector
    • Sample value: Maximum 2 mL, 6 × 25 = 150 units
    • Sample volume: 0 - 1mL, adjustable
    • Sample residual: < 0.02 %
    Column oven
    • Temperature control: Room temperature ± 5 ºC - 80 ºC
    • Constant temperature accuracy: ± 1 ºC
    • Constant temperature precision: ≤ 0.1 ºC
    • Temperature setting resolution ratio: 0.1 ºC
    Overall Instrument
    • Quantitative analysis: RDS 6 ≤ 0.2 %
    • Quantitative repeatability: RDS 6 ≤ 0.2 %
    • Sample residual: ≤ 0.01 %
    • Gradient error: ≤ 0.01 %
    • Inclusion of high performance patent gradient mixer
    • Intelligent anti-control software with powerful control function
    • Stand-alone TFT colour screen with touch control
    • High pressure gradient pump
  • This gradient HPLC system has multiple applications in medicine analysis, detection of protein content of food, environmental analysis, oil and chemical analysis and in scientific research studies.

HPLC system LHLC-B11

  • High pressure constant-current infusion pumpHigh pressure pump
  • Micro volume series dual-pistonInfusion Way
  • 0 - 9999 PsiInfusion pressure
  • ≤ 0.5 %Flow rate

HPLC system LHLC-A11

  • Two sets of pumpHigh pressure pump
  • 0.001 - 9.999 mL / min (0.001 mL / min increments)Flow rate range
  • ≤ 0.5 MPa (0 - 42 MPa)Pressure range
  • RSD < 0.06 %Flow rate precision

HPLC system LHLC-A10

  • Single pumpHigh pressure pump
  • 0.001 - 9.999 mL / min (0.001 mL / min increment)Flow rate range
  • ≤ 0.5 MPa (0 - 42 MPa)Pressure range
  • RSD < 0.06 %Flow rate precision

Analytical HPLC-A7832

  • Model No HPD-A20 HPD-A21
    Wavelength range Single wavelength ( 190 to 800 nm ) Dual wavelength ( 190 to 800 nm )
    Wavelength accuracy ± 0.5 nm ± 2 nm
    Wavelength repeatability 0.15 nm 0.3 nm
    Band width 8 nm 8 nm
    Optical path 10 nm 10 nm
    Deuterium lamp 190 to 400 nm 190 to 400 nm
    Tungsten lamp 400 to 800 nm 400 to 800 nm
    Baseline noise ( Static ) Au ≤ ± 0.35 × 10-5 ≤ ± 0.5 × 10-5
    Baseline drift ( Static ) Au / h ≤ 1 × 10-4 ≤ 1.5 × 10-4
    Minimum detection limit ≤ 4 × 10-8 ≤ 4 × 10-8
    Power 100 W 100 W
    Power supply 90 V / 50 to 60 Hz 90 V / 50 to 60 Hz
    Dimension 460 × 315 × 180 mm 460 × 315 × 180 mm
    Weight 9.5 kg 9.5 kg
    Detector Specifications
  • {"Flow rate range":"0.0001 to 10 ml \/ min","Maximum pressure":"50 MPa , 0 to 2 ml \/ min","Column size":"4.6 \u00d7 250 mm","Sample size":"5 \u00b5m","Pressure pulsation":"\u2264 1 % at 10 MPa","Flow rate accuracy":"\u00b1 0.15 % at 1 ml \/ min","Flow rate repeatability %":"\u2264 0.1","Gradient accuracy %":"\u00b1 1.5","Gradient repeatability %":"\u2264 0.2","Dimension":"460 \u00d7 310 \u00d7 180 mm","Power supply":"90 V","Weight":"11.6 kg"}Pump Specifications

Analytical HPLC-A7840

  • {"Solvent delivery":"Double plunger series mode, floating plunger design, single stroke volume 30 \u00b5l","Pump Head":"10 ml stainless steel pump head","Flow rate range":"0.001 to 9.999 ml \/ min","Flow rate increment":"0.001 ml\/min","Maximum pressure":"42 MPa","Column size":"4.6 \u00d7 250 mm 5 \u00b5m","Pipeline Size":"1\/16 inch","Flow rate accuracy":"\u00b1 0.5 %","Flow rate precision":"RSD \u2264 0.1%","Gradient accuracy %":"\u00b1 0.5 %","Gradient repeatability %":"RSD \u2264 0.15%","GLP":"Automatic recording of pump sealing ring working time","Display":"LCD backlight display","Power":"80 W","Power supply":"220 V \u00b1 10%, 50 Hz","Dimension":"398 \u00d7 149 \u00d7 267 mm"}Pump Specifications
  • {"Wavelength range":"190 to 700 nm","Wavelength accuracy":"\u00b1 1 nm","Wavelength repeatability":"0.2 nm","Deuterium lamp":"190 to 400 nm","Tungsten lamp":"400 to 700 nm","Baseline noise ( Static ) Au":"1 \u00d7 10-5<\/sup> AU\/h at 254 nm with methanol\/water= 80\/20","Baseline drift ( Static ) Au \/ h":"1.5 \u00d7 10-4<\/sup> AU\/h at 254 nm with methanol\/ water= 80\/20","Time constants":"16 grades adjustable","Display":"LCD backlight display","GLP":"Total number of lights, lamp replacement time and product serial number","Power":"75 W","Power supply":"220 V \u00b1 10%, 50 Hz","Dimension (L \u00d7 W \u00d7 H)":"398 \u00d7 149 \u00d7 267 mm"}Detector Specifications
Analytical HPLC-A7835
  • {"Model No":"HPP-B1","Flow rate range":"0.0001 to 10 ml \/ min","Maximum pressure":"42 MPa , 0 to 5 ml \/ min","Column size":"4.6 \u00d7 250 mm","Sample size":"5 \u03bcm","Pressure pulsation":"\u2264 1.5 % at 10 MPa","Flow rate accuracy":"\u00b1 0.2 % at 1 ml \/ min","Flow rate repeatability %":"\u2264 0.15","Gradient accuracy %":"\u00b1 2","Gradient repeatability %":"\u2264 0.2","Dimension":"450 \u00d7 260 \u00d7 190 mm","Power supply":"90 V to 240 V","Weight":"11.2 kg"}Pump Specifications
  • {"Model No":"HPD-B2","Wavelength range":"190 to 800 nm","Wavelength accuracy":"\u00b1 0.75 nm","Wavelength repeatability":"0.2 nm","Band width":"8 nm","Optical path":"10 nm","Deuterium lamp":"190 to 400 nm","Tungsten lamp":"400 to 800 nm","Baseline noise ( Static ) Au":"\u2264 \u00b1 0.5 \u00d7 10 -5<\/sup>","Baseline drift ( Static ) Au \/ h":"\u2264 1.5 x 10 -4<\/sup>","Minimum detection limit":"\u2264 4 x 10 -8<\/sup>","Power":"100 W","Power supply":"90 V \/ 50 to 60 Hz","Dimension":"460 \u00d7 260 \u00d7 190 mm","Weight":"9.2 kg"}Detector Specifications
HPLC system LHLC-A10
  • Single pumpHigh pressure pump
  • 0.001 - 9.999 mL / min (0.001 mL / min increment)Flow rate range
  • ≤ 0.5 MPa (0 - 42 MPa)Pressure range
  • RSD < 0.06 %Flow rate precision

HPLC system LHLC-A11

  • Two sets of pumpHigh pressure pump
  • 0.001 - 9.999 mL / min (0.001 mL / min increments)Flow rate range
  • ≤ 0.5 MPa (0 - 42 MPa)Pressure range
  • RSD < 0.06 %Flow rate precision

HPLC system LHLC-B11

  • High pressure constant-current infusion pumpHigh pressure pump
  • Micro volume series dual-pistonInfusion Way
  • 0 - 9999 PsiInfusion pressure
  • ≤ 0.5 %Flow rate