HPLC system LHLC-A10

This HPLC system is equipped with high-pressure constant-current pump and UV detector. Pressure pulsation is controlled by diaphragmatic damper in seconds. The UV detectors helps in identification of samples and elutes that are colourless, odourless in nature. There is auto boot detection function available in this HPLC system.

  • High pressure pump Single pump
    Flow rate range 0.001 - 9.999 mL / min (0.001 mL / min increment)
    Pressure range ≤ 0.5 MPa (0 - 42 MPa)
    Flow rate precision RSD < 0.06 %
    Pulsation ≤ 0.1 MPa (flow rate 1 ml / min, Pressure 5 - 10 MPa)
    Qualitative analysis RSD 6 ≤ 0.1 % (naphthalene methyl alcohol solvent)
    Quantitative repeatability RSD 6 ≤ 0.3 % (naphthalene methyl alcohol solvent)
    Flow rate setting error Ss ≤ 2 % (1 mL / min water, 5 - 10 MPa, room temperature)
    Air tightness of the pump
    Mixing accuracy ± 1 %
    Mixing precision ± 0.2 %
    Size (W × H × D) 450 × 300 × 160 mm
    UV Detector
    • Wavelength range: 190 - 680 nm
    • Flow cell: 8 µL
    • Spectrum band width: 8 nm
    • Baseline noise: ≤ ± 0.25 × 10<sup>-5</sup> AU (empty cell)
    • Wavelength accuracy: ± 1 nm
    • Baseline drift: ≤ ± 0.4 × 10<sup>-4</sup> AU (empty cell)
    • Wavelength precision: < 0.1 nm
    • Limit of detection: 3 × 10<sup>-9</sup> g / mL (naphthalene methyl alcohol solvent)
    • Size (W × H × D): 450 × 300 × 160 mm
    • Limit baseline noise can be ensured using electronic pulse suppression technique
    • High pressure single pump
    • Easy to change gasket ring & automatic washing function for the column
    • User-friendly software for smooth operations
  • The HPLC has applications in separation of mixture of compounds in chemical analysis, poison analysis, environmental analysis, protein food detection, medicine testing, and polymer analysis.

HPLC system LHLC-A11

  • Two sets of pumpHigh pressure pump
  • 0.001 - 9.999 mL / min (0.001 mL / min increments)Flow rate range
  • ≤ 0.5 MPa (0 - 42 MPa)Pressure range
  • RSD < 0.06 %Flow rate precision

HPLC system LHLC-B11

  • High pressure constant-current infusion pumpHigh pressure pump
  • Micro volume series dual-pistonInfusion Way
  • 0 - 9999 PsiInfusion pressure
  • ≤ 0.5 %Flow rate

HPLC system LHLC-B10

  • High pressure gradient pumpHigh pressure pump
  • 45 MPa, Stainless steel pumpPeak operating pressure
  • ± 0.02 MPaPressure
  • 0.001 - 9.999 mL / minFlow rate range

Analytical HPLC-A7832

  • Model No HPD-A20 HPD-A21
    Wavelength range Single wavelength ( 190 to 800 nm ) Dual wavelength ( 190 to 800 nm )
    Wavelength accuracy ± 0.5 nm ± 2 nm
    Wavelength repeatability 0.15 nm 0.3 nm
    Band width 8 nm 8 nm
    Optical path 10 nm 10 nm
    Deuterium lamp 190 to 400 nm 190 to 400 nm
    Tungsten lamp 400 to 800 nm 400 to 800 nm
    Baseline noise ( Static ) Au ≤ ± 0.35 × 10-5 ≤ ± 0.5 × 10-5
    Baseline drift ( Static ) Au / h ≤ 1 × 10-4 ≤ 1.5 × 10-4
    Minimum detection limit ≤ 4 × 10-8 ≤ 4 × 10-8
    Power 100 W 100 W
    Power supply 90 V / 50 to 60 Hz 90 V / 50 to 60 Hz
    Dimension 460 × 315 × 180 mm 460 × 315 × 180 mm
    Weight 9.5 kg 9.5 kg
    Detector Specifications
  • {"Flow rate range":"0.0001 to 10 ml \/ min","Maximum pressure":"50 MPa , 0 to 2 ml \/ min","Column size":"4.6 \u00d7 250 mm","Sample size":"5 \u00b5m","Pressure pulsation":"\u2264 1 % at 10 MPa","Flow rate accuracy":"\u00b1 0.15 % at 1 ml \/ min","Flow rate repeatability %":"\u2264 0.1","Gradient accuracy %":"\u00b1 1.5","Gradient repeatability %":"\u2264 0.2","Dimension":"460 \u00d7 310 \u00d7 180 mm","Power supply":"90 V","Weight":"11.6 kg"}Pump Specifications

Analytical HPLC-A7835

  • {"Model No":"HPP-B1","Flow rate range":"0.0001 to 10 ml \/ min","Maximum pressure":"42 MPa , 0 to 5 ml \/ min","Column size":"4.6 \u00d7 250 mm","Sample size":"5 \u03bcm","Pressure pulsation":"\u2264 1.5 % at 10 MPa","Flow rate accuracy":"\u00b1 0.2 % at 1 ml \/ min","Flow rate repeatability %":"\u2264 0.15","Gradient accuracy %":"\u00b1 2","Gradient repeatability %":"\u2264 0.2","Dimension":"450 \u00d7 260 \u00d7 190 mm","Power supply":"90 V to 240 V","Weight":"11.2 kg"}Pump Specifications
  • {"Model No":"HPD-B2","Wavelength range":"190 to 800 nm","Wavelength accuracy":"\u00b1 0.75 nm","Wavelength repeatability":"0.2 nm","Band width":"8 nm","Optical path":"10 nm","Deuterium lamp":"190 to 400 nm","Tungsten lamp":"400 to 800 nm","Baseline noise ( Static ) Au":"\u2264 \u00b1 0.5 \u00d7 10 -5<\/sup>","Baseline drift ( Static ) Au \/ h":"\u2264 1.5 x 10 -4<\/sup>","Minimum detection limit":"\u2264 4 x 10 -8<\/sup>","Power":"100 W","Power supply":"90 V \/ 50 to 60 Hz","Dimension":"460 \u00d7 260 \u00d7 190 mm","Weight":"9.2 kg"}Detector Specifications
Analytical HPLC-A7840
  • {"Solvent delivery":"Double plunger series mode, floating plunger design, single stroke volume 30 \u00b5l","Pump Head":"10 ml stainless steel pump head","Flow rate range":"0.001 to 9.999 ml \/ min","Flow rate increment":"0.001 ml\/min","Maximum pressure":"42 MPa","Column size":"4.6 \u00d7 250 mm 5 \u00b5m","Pipeline Size":"1\/16 inch","Flow rate accuracy":"\u00b1 0.5 %","Flow rate precision":"RSD \u2264 0.1%","Gradient accuracy %":"\u00b1 0.5 %","Gradient repeatability %":"RSD \u2264 0.15%","GLP":"Automatic recording of pump sealing ring working time","Display":"LCD backlight display","Power":"80 W","Power supply":"220 V \u00b1 10%, 50 Hz","Dimension":"398 \u00d7 149 \u00d7 267 mm"}Pump Specifications
  • {"Wavelength range":"190 to 700 nm","Wavelength accuracy":"\u00b1 1 nm","Wavelength repeatability":"0.2 nm","Deuterium lamp":"190 to 400 nm","Tungsten lamp":"400 to 700 nm","Baseline noise ( Static ) Au":"1 \u00d7 10-5<\/sup> AU\/h at 254 nm with methanol\/water= 80\/20","Baseline drift ( Static ) Au \/ h":"1.5 \u00d7 10-4<\/sup> AU\/h at 254 nm with methanol\/ water= 80\/20","Time constants":"16 grades adjustable","Display":"LCD backlight display","GLP":"Total number of lights, lamp replacement time and product serial number","Power":"75 W","Power supply":"220 V \u00b1 10%, 50 Hz","Dimension (L \u00d7 W \u00d7 H)":"398 \u00d7 149 \u00d7 267 mm"}Detector Specifications
HPLC system LHLC-B11
  • High pressure constant-current infusion pumpHigh pressure pump
  • Micro volume series dual-pistonInfusion Way
  • 0 - 9999 PsiInfusion pressure
  • ≤ 0.5 %Flow rate

HPLC system LHLC-A10

  • Single pumpHigh pressure pump
  • 0.001 - 9.999 mL / min (0.001 mL / min increment)Flow rate range
  • ≤ 0.5 MPa (0 - 42 MPa)Pressure range
  • RSD < 0.06 %Flow rate precision

HPLC system LHLC-A11

  • Two sets of pumpHigh pressure pump
  • 0.001 - 9.999 mL / min (0.001 mL / min increments)Flow rate range
  • ≤ 0.5 MPa (0 - 42 MPa)Pressure range
  • RSD < 0.06 %Flow rate precision