Gas Chromatography LGC-B11

Gas Chromatography LGC-B11 is a benchtop unit with advanced flow controller (AFC) technology. System has digital control mode and polar capillary columns to separate polar analytes. GC oven is capable of 9 temperature program ramps and hold steps, and features rapid cooling: 250 °C to 50 °C < 10 minutes. Accurate flow rate control via AFC has higher level repeatability of retention time and peak area, enabling specific level of analysis by means of dual FID detectors.

  • Oven
    • Maximum capacity: 22 L
    • Temperature range: RT + 5 ℃ to 400 ℃
    • Temperature accuracy: ± 0.1 ℃
    • Temperature program: 9-phase/ 10 platform
    • Program total time: 9999.9 min
    • Maximum temperature heating rate: 0.1 to 60 ℃/ min
    • Maximum cooling rate: ˂ 10 mins (250 to 50 ℃)
    Sample injector
    • Temperature range: RT+ 7 ℃ to 420 ℃
    • Temperature control: Automatic
    • Injector type: Column or split
    • Maximum sample inlet: 3 pcs
    • Carrier gas flow type: Constant pressure
    • Pre column pressure range: 0 to 400 kPa
    • Pre column pressure accuracy: 0.1 kPa
    • Mobile phase flow range: H2: 0 to 200 ml/min, N2: 0 to 150 ml/min
    Detector
    • Temperature range: RT + 7 ℃ to 420 ℃
    • Maximum installation: Dual detector
    • Flame ignition mode: Automatic
    • Flame Ionization Detector (FID): Logarithmic amplifier <br/> High voltage switch control <br/> Baseline signal display <br/> Ignition coil control <br/> Detection limit: ≤3 ×10 to 12 g/s (Sample: n-hexadecane) <br/> Baseline noise: ≤ 5 × 10 <sup> - 14 </sup> A <br/> Baseline draft: ≤ 6 × 10 <sup> -13 </sup> A <br/> Dynamic range: 10 <sup> 7 </sup> <br/> RSD: ≤ 3 %
    Power supply AC 220 V ± 22 V, 50 Hz ± 0.5 Hz, 3000 W
    Packing dimension 1020 × 660 × 740 mm
    Gross Weight 82 kg
    Net weight 70 kg
    • 7-inch LCD display specifies temperature, gas flow and heating curve
    • H2 and N2 mobile gas phase with AFC and constant pressure gas flow type
    • Equipped with capillary analysis system and capillary split sampler
    • Injection system incorporates diaphragm cleaning, split flow and auxiliary gas adjustment
    • Standard unit for high-speed analysis with polar capillary column
    • PEG-20M and FFAP stationary phase polar capillary columns can be coupled
    • Dual FID detector with automatic flame ignition and baseline signal display
    • Self-diagnostic and automatic fault recognition function
    • Multi-core and 32-bit embedded hardware system ensures the reliable operation
    • Equipped with chromatography workstation software to process data
    • Memory function can store 20 sample test data
    • RS232 interface and LAN port
  • Suitable for analysis of trace toxic compounds, monitoring and research in environmental protection, atmosphere and water source pollution. It is widely used for petrol-chemistry, environmental protection, epidemic prevention, pharmacology and scientific research.

  • Accessories no Accessories name
    1.        FID detector system
    2.        Capillary injection system
    3.        Chromatography workstation software
  • Accessories no Accessories name
    1 Polar capillary columns PEG-20M
    Stationary phase: 30 m × 0.32 mm
    injector temperature: 250 ℃
    Oven temperature: 165 ℃
    Detector: FID- 250 ℃
    Inlet pressure: 0.04 Mpa
    2 Polar capillary columns FFAP
    Stationary phase: 30 m × 0.32 mm
    Injector temperature: 250 ℃
    Oven temperature: 165 ℃
    Detector: FID- 250 ℃
    Inlet pressure: 0.04 Mpa
    3 Syringe 1 ul
    4 Syringe 5 ul

Ion Chromatograph LPIC-A10

  • {"Type":"High-pressure and low \u2013pulse double plunger tandem advection pump","Maximum Pressure":"35 MPa (PEEK)","Flow Range":"0.001 to 9.999 mL\/min","Flow accuracy":"\u00b1 0.5 %","Flow Repeatability":"RSD \u2264 0.1%","Flow stability":"(0.2 to 0.5) mL\/min, \u2264 3 %\n(0.5 to 1.0) mL\/min, \u2264 2 %\n>1.0 mL\/min, \u2264 2 %"}Chromatographic Pump
  • {"Temperature range":"Room temperature +5\u00b0C to 60\u00b0C (41 to 140\u00b0F)","Allowable deviation":"\u00b1 1\u00b0C","Temperature stability":"\u2264 0.5\u00b0C\/h"}Built-in Heat conduction type column heater
  • {"Maximum pressure":"35 MPa","Control mode":"By stepper motor","Power supply":"24 V (DC)"}Electromagnetic injection valve
  • {"Type":"Temperature control and bipolar conductivity detector","Detection type":"Bipolar conductivity detection","Cell volume":"\u2264 0.8 \u00b5L","Measurement range":"0 to 45000 \u00b5 S\/cm (adjustable)","Detection resolution":"\u2264 0.0020 nS\/cm","Output voltage":"-6000 to 6000 mV (adjustable)","Temperature range":"Room temperature +5\u00b0C to 60\u00b0C (41 to 140\u00b0F)","Temperature accuracy":"\u00b1 0.01\u00b0C","Maximum pressure":"10 MPa","Baseline drift":"\u2264 20 % FS\/30 min","Baseline noise":"\u22640.5 % FS","Instrument linearity":"\u2265 0.999","Quantitative Repeatability":"\u22640.5 %","Qualitative Repeatability":"\u22642 %","Minimum detectable concentration":"Cl-\u22640.005 \u00b5g\/mL\nLi+ \u22640.005 \u00b5g\/mL"}Digital and temperature control detection system

Ion chromatograph system LICS-A20

  • {"Type":"High pressure and low pulse double plunger tandem advection pump","Maximum pressure":"35 MPa (PEEK)","Pressure display accuracy":"\u2264 0.1 MPa","Flow stability":"(0.2 to 0.5) ml\/ min \u2264 3 %\n(0.5 to 1.0) ml\/ min \u2264 2 %\n> 1.0 ml\/ min 2 %","Flow range":"0.001 to 9.999 ml\/ min","Pressure pulse":"\u2264 0.5%"}Ion Chromatographic Pump
  • {"Type":"Temperature Control and Bipolar Conductivity Detector","Detection mode":"Bipolar conductivity detection","Cell volume":"\u2264 0.8 \u00b5L","Detection range":"0 to 35000 \u00b5S\/ cm","Detection resolution":"\u2264 0.0020 nS\/ cm","Output voltage":"- 6000 to +6000 mV (adjustable)","Baseline noise":"\u2264 0.05 % F.S.","Baseline drift":"\u2264 3 % FS\/ 30 min","Operating temperature range":"Room temperature +5 to 60\u00b0C (41 to 140\u00b0F)","Controlling temperature accuracy":"\u00b1 0.01\u00b0C","Maximum pressure":"10 MPa","Linear range":"\u2265 103","Instruments linearity":"\u2265 0.999","Quantitative repeatability":"\u2264 1.0 %","Qualitative repeatability":"\u2264 1.0 %","Minimum detectable concentration":"Cl \u2264 0.0005 ug\/ ml\nBrO3 \u2264 0.001 ug\/ ml\nLi+ \u2264 0.001 ug\/ ml"}Conductivity detector
  • {"Plastic flow path":"Made of PEEK materials","Six way Valve":"PEEK material, 5000 psi pressure, independent automatic collection and flow function"}Flow system
  • {"Inner volume":"300 \u00b5l","Degassing volume":"7.5 ml","Maximum flow rate":"10 ml\/ min","Degassing efficiency":"Degassing 90 % at 1ml\/ min","Organic tolerance":"1"}Degassing Device
Ion chromatograph system LICS-A22
  • {"Type":"High pressure and low pulse double plunger tandem advection pump","Maximum pressure":"35 MPa (PEEK)","Display pressure accuracy":"\u2264 0.1 MPa","Flow stability":"(0.2 to 0.5) ml\/ min \u2264 3 % , (0.5 to 1.0) ml\/ min \u2264 2 %, > 1.0 ml\/ min \u2264 2 %","Flow range":"0.001 to 9.999 ml\/ min","Pressure pulse":"\u2264 0.5 %"}Ion Chromatographic pump
  • {"Maximum pressure":"35 MPa","Contact material of the rotor":"PEEK","Control mode":"By Stepper motor","Power supply":"DC 24 V"}Numerical control and electromagnetic injection
  • {"Operating temperature range":"Room temperature 20 to 60\u00b0C ( 68 to 140\u00b0F)","Controlling temperature accuracy":"\u00b1 0.01\u00b0C","Allowable deviation of column heater temperature":"\u00b1 2\u00b0C","Temperature stability":"\u2264 1\u00b0C\/ h"}Column heater
  • {"Plastic flow path":"Made up of PEEK material","Six way valve":"PEEK material pressure 5000 psi, independent automatic collection and flow function"}Flow system
Ion chromatograph system LICS-A21
  • {"Type":"High pressure and low pulse double plunger tandem advection pump","Maximum pressure":"35 MPa (PEEK)","Pressure display accuracy":"\u2264 0.1 MPa","Flow stability":"(0.2 to 0.5) ml\/ min \u2264 3 %\n(0.5 to 1.0) ml\/ min \u2264 2 %\n> 1.0 ml\/ min 2 %","Flow range":"0.001 to 9.999 ml\/ min","Pressure pulse":"\u2264 0.5%"}Ion Chromatographic Pump
  • {"Type":"Temperature Control and Bipolar Conductivity Detector","Detection mode":"Bipolar conductivity detection","Cell volume":"\u2264 0.8 \u00b5L","Detection range":"0 to 45000 \u00b5S\/ cm","Detection resolution":"\u2264 0.0020 nS\/ cm","Output voltage":"- 6000 to +6000 mV (adjustable)","Baseline noise":"\u2264 0.05 % F.S.","Baseline drift":"\u2264 3 % FS\/ 30 min","Operating temperature range":"Room temperature +5 to 60\u00b0C (41 to 140\u00b0F)","Controlling temperature accuracy":"\u00b1 0.01\u00b0C","Maximum pressure":"10 MPa","Linear range":"\u2265 103","Instruments linearity":"\u2265 0.999","Quantitative repeatability":"\u2264 1.0 %","Qualitative repeatability":"\u2264 1.0 %","Minimum detectable concentration":"Cl \u2264 0.0005 ug\/ ml\nBrO3 \u2264 0.001 ug\/ ml\nLi+ \u2264 0.001 ug\/ ml"}Conductivity detector
  • {"Plastic flow path":"Made of PEEK materials","Six way Valve":"PEEK material, 5000 psi pressure, independent automatic collection and flow function"}Flow system
  • 150 WPower supply

Gas Chromatography LGC-B11

  • {"Maximum capacity":"22 L","Temperature range":"RT + 5 \u2103 to 400 \u2103","Temperature accuracy":"\u00b1 0.1 \u2103","Temperature program":"9-phase\/ 10 platform","Program total time":"9999.9 min","Maximum temperature heating rate":"0.1 to 60 \u2103\/ min","Maximum cooling rate":"\u02c2 10 mins (250 to 50 \u2103)"}Oven
  • {"Temperature range":"RT+ 7 \u2103 to 420 \u2103","Temperature control":"Automatic","Injector type":"Column or split","Maximum sample inlet":"3 pcs","Carrier gas flow type":"Constant pressure","Pre column pressure range":"0 to 400 kPa","Pre column pressure accuracy":"0.1 kPa","Mobile phase flow range":"H2: 0 to 200 ml\/min, N2: 0 to 150 ml\/min"}Sample injector
  • {"Temperature range":"RT + 7 \u2103 to 420 \u2103","Maximum installation":"Dual detector","Flame ignition mode":"Automatic","Flame Ionization Detector (FID)":"Logarithmic amplifier \nHigh voltage switch control \nBaseline signal display \nIgnition coil control \nDetection limit: \u22643 \u00d710 to 12 g\/s \n(Sample: n-hexadecane) \nBaseline noise: \u2264 5 \u00d7 10 - 14 <\/sup> A \nBaseline draft: \u2264 6 \u00d7 10 -13 <\/sup> A \nDynamic range: 10 7 <\/sup> \nRSD: \u2264 3 %"}Detector
  • AC 220 V ± 22 V, 50 Hz ± 0.5 Hz, 3000 WPower supply

Gas chromatography mass spectrometry GC-MS-879

  • {"Model No":"GC-87","Flow rate":"1 ~ 1.5 ml \/ min","Maximum flow rate":"10 ml \/ min","Inlet temperature":"450 \u00b0C","Pressure range":"0 ~ 100 psi","Heating rate":"Upto 120 \u00b0C \/ min","Room temperature":"4 \u00b0C ~ 450 \u00b0C","Pressure control mode":"Electronic pressure control ( Supports CV and CC )","Split mode":"Split \/ split less","Split ratio":"1000 : 1","Temperature programming":"7 stages \/ 8 platforms","Auto sampler":"Optional"}Gas Chromatography Specifications
  • {"Model No":"MS-89","Mass range":"1.5 ~ 1000 amu","Ion \u2013 source temperature":"100 \u00b0C ~ 350 \u00b0C","GC-MS interface temperature":"450 \u00b0C","Scan rate":"Upto 10000 amu\/s","Stability":"\u00b1 0.10 amu \/ 48 hrs.","Filament emission current":"0 ~ 350 \u00b5A","EI source ionization energy":"5 eV ~ 250 eV","Resolution":"Unit resolution","Detector":"High energy dynode electron multiplier","Sensitivity":"Full scan ( S \/ N is \u2265 30 : 1 )","Vacuum":"Turbo molecular pump ( 67 L\/s )"}Mass Spectrometry Specifications