\nHigh voltage switch control \nBaseline signal display \nIgnition coil control \nDetection limit: \u22643 \u00d710 to 12 g\/s \n(Sample: n-hexadecane) \nBaseline noise: \u2264 5 \u00d7 10 - 14 <\/sup> A \nBaseline draft: \u2264 6 \u00d7 10 -13 <\/sup> A \nDynamic range: 10 7 <\/sup> \nRSD: \u2264 3 %"}. Explore the product to find out more. Visit Labtron.best!"> Gas Chromatography LGC-B11 Catalog | Labtron \nHigh voltage switch control \nBaseline signal display \nIgnition coil control \nDetection limit: \u22643 \u00d710 to 12 g\/s \n(Sample: n-hexadecane) \nBaseline noise: \u2264 5 \u00d7 10 - 14 <\/sup> A \nBaseline draft: \u2264 6 \u00d7 10 -13 <\/sup> A \nDynamic range: 10 7 <\/sup> \nRSD: \u2264 3 %"}. Explore the product to find out more. Visit Labtron.best!"> \nHigh voltage switch control \nBaseline signal display \nIgnition coil control \nDetection limit: \u22643 \u00d710 to 12 g\/s \n(Sample: n-hexadecane) \nBaseline noise: \u2264 5 \u00d7 10 - 14 <\/sup> A \nBaseline draft: \u2264 6 \u00d7 10 -13 <\/sup> A \nDynamic range: 10 7 <\/sup> \nRSD: \u2264 3 %"}. Explore the product to find out more. Visit Labtron.best!">

Gas Chromatography LGC-B11 Catalog

Overview

Gas Chromatography LGC-B11 is a benchtop unit with advanced flow controller (AFC) technology. System has digital control mode and polar capillary columns to separate polar analytes. GC oven is capable of 9 temperature program ramps and hold steps, and features rapid cooling: 250 °C to 50 °C < 10 minutes. Accurate flow rate control via AFC has higher level repeatability of retention time and peak area, enabling specific level of analysis by means of dual FID detectors.

Specifications
Oven
  • Maximum capacity: 22 L
  • Temperature range: RT + 5 ℃ to 400 ℃
  • Temperature accuracy: ± 0.1 ℃
  • Temperature program: 9-phase/ 10 platform
  • Program total time: 9999.9 min
  • Maximum temperature heating rate: 0.1 to 60 ℃/ min
  • Maximum cooling rate: ˂ 10 mins (250 to 50 ℃)
Sample injector
  • Temperature range: RT+ 7 ℃ to 420 ℃
  • Temperature control: Automatic
  • Injector type: Column or split
  • Maximum sample inlet: 3 pcs
  • Carrier gas flow type: Constant pressure
  • Pre column pressure range: 0 to 400 kPa
  • Pre column pressure accuracy: 0.1 kPa
  • Mobile phase flow range: H2: 0 to 200 ml/min, N2: 0 to 150 ml/min
Detector
  • Temperature range: RT + 7 ℃ to 420 ℃
  • Maximum installation: Dual detector
  • Flame ignition mode: Automatic
  • Flame Ionization Detector (FID): Logarithmic amplifier <br/> High voltage switch control <br/> Baseline signal display <br/> Ignition coil control <br/> Detection limit: ≤3 ×10 to 12 g/s (Sample: n-hexadecane) <br/> Baseline noise: ≤ 5 × 10 <sup> - 14 </sup> A <br/> Baseline draft: ≤ 6 × 10 <sup> -13 </sup> A <br/> Dynamic range: 10 <sup> 7 </sup> <br/> RSD: ≤ 3 %
Power supply AC 220 V ± 22 V, 50 Hz ± 0.5 Hz, 3000 W
Packing dimension 1020 × 660 × 740 mm
Gross Weight 82 kg
Net weight 70 kg

Features

  • 7-inch LCD display specifies temperature, gas flow and heating curve
  • H2 and N2 mobile gas phase with AFC and constant pressure gas flow type
  • Equipped with capillary analysis system and capillary split sampler
  • Injection system incorporates diaphragm cleaning, split flow and auxiliary gas adjustment
  • Standard unit for high-speed analysis with polar capillary column
  • PEG-20M and FFAP stationary phase polar capillary columns can be coupled
  • Dual FID detector with automatic flame ignition and baseline signal display
  • Self-diagnostic and automatic fault recognition function
  • Multi-core and 32-bit embedded hardware system ensures the reliable operation
  • Equipped with chromatography workstation software to process data
  • Memory function can store 20 sample test data
  • RS232 interface and LAN port
Standard Accessories :

Accessories no Accessories name
1.        FID detector system
2.        Capillary injection system
3.        Chromatography workstation software

Optional Accessories :

Accessories no Accessories name
1 Polar capillary columns PEG-20M
Stationary phase: 30 m × 0.32 mm
injector temperature: 250 ℃
Oven temperature: 165 ℃
Detector: FID- 250 ℃
Inlet pressure: 0.04 Mpa
2 Polar capillary columns FFAP
Stationary phase: 30 m × 0.32 mm
Injector temperature: 250 ℃
Oven temperature: 165 ℃
Detector: FID- 250 ℃
Inlet pressure: 0.04 Mpa
3 Syringe 1 ul
4 Syringe 5 ul

Applications :