Touch Thermal Cycler (Basic) LTCB-A13 Catalog

Overview

Touch Thermal Cycler (Basic) LTCB-A13 with customized 6 pcs long service life peltier heating units. With 5°C/s maximum Ramping rate, this cycler saves operation time of user. It has 3 × (32 × 0.2 ml) capacity and temperature ranges from 0 to 100℃. Designed with 8 inch TFT touch screen with graphical real time display. Features with PC control function, print function, large storage capacity, USB connection, this cycler meets latest experimental demands. Adopts 3 blocks independently controlled and run 3 different PCR programs simultaneously. It is used in advanced PCR applications across research and development etc.

Specifications
Capacity 3 × (32 × 0.2 ml )
Temperature Range 0 to 100℃
Heating Rate 5 ℃/s
Cooling Rate 4 ℃/s
Uniformity ≤± 0.3℃
Accuracy ≤± 0.2℃
Display Resolution 0.1℃
Temperature Control Block\Tube
Ramping Rate Adjustable 0.1 to 5℃
Hot Lid Temperature 30 to 110℃
Hot Lid Height Adjustable Step less Adjustable
Number of Programs 10000 + (USB FLASH)
Max. No. of Step 30
Max. No. of Cycle 100
Time Increment / Decrement 1 Sec to 600 Sec
Temperature Increment / Decrement 0.1 to 10.0℃
Pause Function Yes
Auto Data Protection Yes
Long PCR function yes
Hold at 4℃ Forever
Language English
Print Yes
LAN to computer Yes
LCD 8 inch, 800 × 600 pels, TFT
Communication USB 2.0, LAN
Dimension 390 × 270 × 255 mm
Power consumption 600 W
Power Supply 85 to 264 VAC , 47 to 63 Hz , 600 W
Weight 9 kg

Features

  • Designed with 6 long service life Peltier heating unit and form 3 circuits to control 3 temperature zones
  • High heating and cooling rate, maximum ramping rate is 5°C/s
  • Reinforced aluminum module with anodizing technology and rapid heating-conducting property, have enough corrosion resistance
  • Designed with 8 inch, 800 × 600 pels, TFT LCD display
  • Step less adjustable hot lid with pressure protection, fit tubes of different heights to melt and avoid tube evaporation
  • Adopts 3 blocks independently controlled and run 3 different PCR programs simultaneously
  • Built-in 11 standard program file template, quickly edit the required files
  • Equipped with one click incubation function such as denaturation enzyme cutting/enzyme-link and ELISA
  • Automatic restart after power failure, when power is restored to continue to run unfinished program
  • GLP report records every step to provide accurate data support for experiment result analysis
  • Offers login management, three-tier permission, password protection function to ensure data security
  • Support USB and LAN to update software
  • Support Email alert function when experimental is over
Applications :

Touch Thermal Cycler is used in used in advanced PCR applications across research and development, animal and plant tissue culture, food science, pharmaceuticals, life science, and analytical laboratories.